Etanercept (Enbrel) Pharmacokinetic ELISA

Etanercept (Enbrel) Pharmacokinetic ELISA

1211.24 EUR

1211.24 EUR
In Stock
quantity
Produktdetaljer

Katalognummer: 1061 - EL-1611-051

Produktkategori: Företag och industri > Vetenskap och laboratorium

Leverantör:

AffinitylmmunoGentaur

Storlek: 1X96 wells

Parameters

Additional informationComponents: Coated microtiter plate, 96 wellsCalibrator diluent. - 1.8mlCalibrator 12ul10X wash buffer - 25mlAssay buffer - 50ml1000X detection reagent - 17ulTMB - 12mlTMB stop solution - 12mlPlate sealers - 3 ; Assay procedure: This assay employs the sandwich enzyme immunoassay technique. Capture antibody is coated onto a 96 well microplate. Calibrator and test samples are pipetted into the appropriate wells. Etanercept present in biological matrices is bound by the immobilized anti-Etanercept antibody. After washing away any unbound substances, enzyme linked anti-Etanercept antibody is added to the wells. The plate is washed to remove any unbound antibody-enzyme reagent and a substrate solution is added to the wells for color development. The color development is proportional to the amount of Etanercept present in test samples. The color development is stopped and the intensity of the color is measured.; Reagent preparation: Prepare only the appropriate amount of required reagent on the day of use. Store all reagents as per instructions stated on the label. 1. Wash Buffer (1X) Preparation: Dilute wash buffer concentrate with ultra-pure water 1/10 before use (for example add 20mL concentrate to 180mL ultra-pure water). Mix well. 2. Detection Reagent (1X) Preparation: Dilute detection reagent with assay buffer 1/1000 before use (for example add 11μl concentrate to 11ml of assay buffer). Mix well. 3. Preparation of Calibrators: Prepare calibrators with concentrations ranging from 2500 ng/mL to 78 ng/mL. The following is an example calibrator curve.; Results calculation: 1. Construct a standard curve by plotting the absorbance obtained from each standard against concentration. Use a 4 or 5 parameter curve fit. Alternatively a log-log curve fit may be used. 2. The concentration of the unknowns can be read directly from this standard curve using the absorbance value for each sample. 3. Any sample undiluted or diluted still reading greater than the highest standard should be diluted appropriately with assay buffer and retested. If the samples have been diluted, the concentration determined from the standard curve must be multiplied by the dilution factor.; Sample collection: This kit is compatible with EDTA-plasma, heparinplasma and serum samples. Samples can be stored at or below -20°C for up to 1 year.; Sample preparation: Dilute calibrators and test samples 1/50 with assay buffer (for example add 5µL of prepared calibrator or sample to 245µL of assay buffer). Mix well. Do not store diluted samples.
Storage and shipping-20C, 1 year
Storage and shippingWith